Sigma Spectrum 6.05.14 wireless B/G

$1,195

The SIGMA Spectrum Infusion System is redefining smart technology with Second Generation software to help prevent pump-related medication errors

SKU: BSP Category: Brand:
No Sales Tax
1 year warranty on equipment orders
Commercial financing available for orders over $2,000

Sigma spectrum 6.05.14 Infusion pump with wireless B/G battery

Model# BAX35700

Certified with 1 year warranty

The Baxter Sigma Spectrum 6.05.14  Infusion Pump comes Patient Ready with AC adapter, pole clamp and wireless B/G  battery. Software version 6.05.14

The SIGMA Spectrum 6.05.14 Infusion System reduces errors in programming IV infusion. The SIGMA Spectrum is easy to use with automatic start-up of the Master Drug Library (MDL) and capability to integrate infusion data into the Electronic Medical Record (EMR).

The SIGMA Gateway provides bi-directional communication capabilities to the SIGMA Spectrum through a hospital’s existing b/g wireless network. This allows delivery of the MDL to an entire fleet of pumps as well as continuous Quality Improvement (CQI) reports to the SIGMA Gateway.

Features:

  • Automatic MDL start-up when machine powers on
  • Automatic drug library updates through wireless searching of updates, downloads and installs
  • Lightweight, compact design promotes patient mobility
  • Bright color screen
  • Lithium-ion battery pack

 

 

Item Name: Spectrum, Baxter calibrated Manufacturer: Sigma International Equipment Category: Infusion Pump Mounting/Use: Pole-Mounted Therapy Type: Large Volume Size: 5.6″ x 4.25″ x 1.75″ Weight: 2.5 lbs., including char

Weight 0.00 lbs
Dimensions 0.00 × 0.00 × 0.00 in

Reviews

There are no reviews yet.

Be the first to review “Sigma Spectrum 6.05.14 wireless B/G”
  • Dimensions:
    • Standard Battery:
      • Without IV Pole Clamp: 5.8″ x 4.2″ x 2.5″ (147 mm x 107 mm x 63.5 mm).
      • With IV Pole Clamp: 5.8″ x 6.4″ x 4.7″ (147 mm x 163 mm x 119 mm).
    • Wireless Battery Module:
      • Without IV Pole Clamp: 6.3″ x 4.2″ x 2.5″ (160 mm x 107 mm x 63.5 mm).
      • With IV Pole Clamp: 6.3″ x 6.4″ x 4.7″ (160 mm x 163 mm x 119 mm).
  • Weight:
    • Standard Battery:
      • Without IV Pole Clamp: 25 oz ± 1 oz (708 grams ± 28 grams).
      • With IV Pole Clamp: 33.5 oz ± 1 oz (950 grams ± 28 grams).
    • Wireless Battery Module: Less than or equal to 0.75 lb (340 grams).
  • Battery:
    • Standard Battery:
      • Lithium Ion, 1800 mA/h, 7.2V nominal. SIGMA Part Number, 35702.
      • Pump operating time on battery power is at least 8 hours at 125 ml/hour with the backlight on (new battery).
      • 12 hour recharge time.
      • Charging occurs if AC Power Adaptor is plugged in whether pump is ON or OFF.
    • Wireless Battery Module:
      • Lithium Ion, 1800 mA/h, 7.2V nominal. SIGMA Part Number, 35083.
      • Capacity 4 hrs (500 mL) at 125 mL/hour at the highest backlight settings.
      • 12 hour recharge time.
      • Charging occurs if AC Power Adaptor is plugged in whether pump is ON or OFF.
  • Pump Characteristics and Flow Rate Accuracy:
    • Linear peristaltic mechanism.
    • 0.5 to 999 mL/hour flow range in 0.1 mL/hour increments from 0.5 to 99.9 mL/hour and 1 mL/hour increments thereafter.
    • Volumetric accuracy (based on volume collected over one hour) using standard Hospira, Baxter.
    • ± 5% from 2 to 800 mL/hour, greater than 800 mL/hour ± 10% for Hospira ± 5% from 2 to 999 mL/hour for Baxter, 0.5 to 1.9 m/hour ± 0.1 mL. Total volume up to 9 liters (Hospira) or 12 liters (Baxter).
    • KVO – Keep vein open rate of either 1 mL/hour or the actual rate, whichever is lower, at infusion complete alarm.
    • Maximum pump pressure – 36 PSI (which can never be obtained with operational occlusion alarm limits).
  • Wireless Network Interface:
    • Standard: IEEE 802.11b.
    • Frequency: 2.4 GHz.
    • Data Rate: Up to 11 Mbps with automatic fallback.
    • Modulation: CCK (11/5 Mbps), DQPSK (2 MBPS), DBPSK (1 MBPS).
    • Transmit Power: 16 dBm typical.
    • Receiver Sensitivity: -82 dBm at 11 Mbps.
  • Wireless Security:
    • WEP (Wired Equivalent Privacy): 64/128-bit encryption (RC4).
    • WPA/WPA2/802.11i:
      • 128-bit TKIP/CCMP encryption.
      • 802.1 x EAP authentication:
        • LEAP (WEP only), PEAP, TTLS.
        • GTC, MD5, OTP, PAP, CHAP, MSCHAP, MSCHAPc2, TTLS-MSCHAPv2.
      • Pre-shared key mode (PSK).
  • Environmental Limits:
    • With Standard Battery:
      • Operating Temperature: 60 to 90 °F (15.6 to 32.3 °C), 20 to 90% relative humidity non-condensing.
      • Storage Temperature: -4 to 120 °F (-20 to 49 °C), 10 to 90% relative humidity non-condensing.
    • With Wireless Battery Module:
      • Operating Temperature: 60 to 80 °F (15.6 to 26.7 °C), 20 to 90% relative humidity non-condensing.
      • Storage Temperature: -4 to 120 °F (-20 to 49 °C), 10 to 90% relative humidity non-condensing.
  • Display: Full color, HRTFT, 240 x 270, LED front-lit, 0.2235 mm x 0.2235 mm dot pitch.
  • Alarms and Alerts:
    • Air-In-Line: Dual beam ultrasonic detector alarms for large bubbles but allows smaller bubbles to pass. Detects air bubbles greater than 1″ (≈ 125 μL Hospira, ≈ 140 μL Baxter), will alarm if greater than 1 mL of air in 15 min., less than 50 μL bubbles are omitted in the summation of the 1 mL.
    • Audio: Speaker actuated audio alarm, low, medium, and high levels selected through the configurations screen.
    • Battery Missing: Pump does not detect battery attached.
    • Depleted Battery: Pump stops running, alarms for 3 minutes.
    • Dose Rate Limit Exceeded: The pump will run with a soft dose rate exceeded after a double confirmation, it will not run with a hard dose rate exceeded, rates must be reset within hard limits.
    • Downstream Occlusion: Automatic restart occurs after the downstream occlusion is cleared. Actuation can be set to Low, 6 ± 4 PSI, Medium, 13 ± 6 PSI, or High, 19 ± 9 PSI.
    • Inactivity: Actuates after the pump has been inactive for 2 minutes.
    • Infusion Complete: Occurs when the VTBI reaches zero, at which time a KVO rate begins (the previous running rate or 1 mL/hour, whichever is lower).
    • In Stop: Load set.
    • In Stop: Open side clamp.
    • In Stop: Push run.
    • Low Battery: Less than 30 minutes of battery power remain.
    • Shut Door.
    • Side Clamp Closed: Pump stopped.
    • System Error.
    • Upstream Occlusion.
    • Very Low Battery: Less than 15 minutes of battery power remain.
  • Timekeeping: Real time clock, battery backed, 10 year life.
  • Logging Memory:
    • 24 hour memory of all set up screens except for ramp/taper modes that are maintained permanently.
    • Separate pump and drug library history logs, minimum of 96 hours each under extensive logging intensive operating conditions.
  • AC Power:
    • AC Power Adaptor, low profile, covers only one outlet, Medical Grade (EN60601-1-2), Input: 100 V-AC to 240 V-AC, 50 to 60 Hz/200 mA, Output: 9 V-DC / 800 mA, short circuit protected, cord length 3.0 m (~ 9.75′). Use only SIGMA part number 55079 or equivalent.
    • The SIGMA Spectrum Infusion Pump is classified according to Medical Electrical Equipment standards as:
      • Class II Equipment.
      • Type BF Applied Part.
      • Continuous Operation.
  • External Interfaces: RS-232 (RTX and CTS only), IrDA (SIR Encoding Protocol. Supports IrOBEX). Additional Asynchronous Serial Port Expansion bus available at battery terminals. Software upgrades may be performed through external RS-232 or IrDA ports.
  • Standards:
    • IEC60601-1 including collateral standards; Third Party Notified Body Testing (Reference Electromagnetic Compatibility Tables).
    • IrDA Serial Infrared Physical Layer Link Specification v1.4(IrPHY), IrDA Serial Infrared Link Access Protocol v1.1 (IrLAP) and IrDA Serial Infrared Link Management Protocol v1.1 (IrLMP), IrDA Tiny TP v1.1.
    • Wireless – 802.11b.

Click here to download the PDF file.